Bioruptor® in Action Video

Newsletter

Subscribe to our newsletter:

Bioruptor® NGS

Back to listing

Description

click here to check recent publications The Bioruptor® NGS is the perfect tool for NGS library preparation. This cutting-edge sonication device is dedicated exclusively to DNA shearing and provides you suitable DNA for unbiased library preparation prior to next generation sequencing. You will obtain your fragmented DNA up to five times faster than with other standard fragmentation devices.

>> Check our current promotions


Bioruptor® NGS sonication device with 0.5/0.65 ml tube holder, Water Cooler & Connector kit UCD-600-TS 1 unit CONTACT US

Bioruptor NGS UCD-600

 

• Simultaneous sonication of 12 samples - short processing time

• Easy-to-use

• Cost effective solution

 

Broad Size Range

Figure 1. Genomic human DNA has been sheared using the Bioruptor® NGS following validated protocols to obtain fragments from 60-1500 bp. All samples were analysed on Bioanalyzer 2100 using DNA High Sensitivity chip.

 

Highly Reproducible and Tight Size Distribution

Figure 2. Six genomic human DNA samples were sheared using the Bioruptor® NGS for 15 minutes. All samples were analysed on a Bioanalyzer 2100 using DNA High Sensitivity chip.

 

Shear DNA 5 x faster than Competitor C

Figure 3. Comparison of processing time (minutes) per sample for different fragment sizes (150 bp, 200 bp, 500 bp, 1000 bp). Processing time includes hands-on-work and sonication time.

 

Bioruptor® NGS System - Overview

The Bioruptor® NGS system consists of: sonication device (water bath and control unit), cooling system and connector  kit for water cooler.

The Connector kit for Bioruptor® Water Cooler controls the water flow between the Bioruptor Twin (UCD-400) and the water cooler. Cold water cycles into the Bioruptor® water tank only during the off cycles (no sonication) and is stopped during the on cycles (sonication). This reduces any water flow disturbance in the tank and allows for more efficient sonication of samples.

 

Validated protocols

Target size
Cycle condition (On/Off cycle)
Cycle number
Total sonication time
150 pb
30 sec/30 sec
30
30 min
175 pb
30 sec/30 sec
20
20 min
200 pb
30 sec/30 sec
13
13 min
250 pb
30 sec/90 sec
7
14 min
350 pb
15 sec/90 sec
9
15 min 45 sec
450 pb
15 sec/90 sec
6
10 min 30 sec
700 pb
15 sec/90 sec
4
7 min
850 pb
15 sec/90 sec
3
5 min 15 sec
1000-1100 pb
5 sec/90 sec
6
9 min 30 sec
1200-1300 pb
5 sec/90 sec
4
6 min 20 sec

 

Faster results and smaller fragments than standard Bioruptors


DNA Shearing (e.g. NGS library preparation)

RNA Shearing

*Check our protocols and product information sections for the corresponding protocols or recommendations of use